Services
NATA Calibration
ACCREDITATION NO: 14283
This laboratory complies with the requirements of ISO/IEC 17025 (1999)
1.41 Pressure and vacuum measuring devices
- Calibration in the range -100 to 0 kPa and 10 kPa to 70 MPa
with least uncertainties of measurement of -
0.2% from -100 to 0 kPa;
0.1% from 10 kPa to 70 MPa;
0.16% for on-site calibration (industrial grade only)
- .01 Pressure gauges
- Including test gauges as defined in AS 1349
.02 Vacuum gauges (bourdon tube)
- Industrial gauges as defined in AS 1349
.11 Pressure transducers
- Self indicating transducers and electrical transducers with dc voltage, dc current or resistance outputs
.12 Pressure recorders
.21 Mercury manometers
- Industrial manometers
.22 Other liquid manometers
- Industrial manometers
Accreditation No: 14283
Corporate Site No: 14277
(Scope Last Changed 30/05/03) |
3.02 Resistors, resistance boxes and potential dividers
- .01 Precision resistors, resistance boxes and conductance boxes
- with least uncertainties of measurement of -
Based on 3.10.09
3.09 Instrument calibrators
- with least uncertainties of measurement of -
Based on 3.10
- .01 D.C. voltage
.02 A.C. voltage
.11 D.C. current
.12 A.C. current
.51 Resistance
3.10 Indicating and recording instruments
- .01 D.C. voltmeters
- with least uncertainties of measurement of -
60 ppm + 1 μV up to 100 mV
60 ppm + 10 μV from 100 mV to 1 V
60 ppm + 100 μV from 1 to 10 V
60 ppm + 1 mV from 10 to 100 V
60 ppm + 10 mV from 100 to 1 000 V
.02 A.C. voltmeters
- At frequencies from 40 40 Hz to 1 kHz
with least uncertainties of measurement of -
0.04% + 10 μV up to 10 mV
0.02% + 10 μV from 10 to 100 mV
0.02% + 100 μV from 100 mV to 10 V
0.03% + 10 mV from 10 to 100 V
0.046% + 100 mV from 100 to 1 000 V
.03 D.C. ammeters
- with least uncertainties of measurement of -
0.044% + 10 nA up to 100 nA
0.013% + 10 nA from 100 nA to 1 μA
12 ppm + 10 nA from 1 to 10 μA
12 ppm + 100 nA from 10 μA to 1 mA
12 ppm + 300 nA from 1 to 10 mA
13 ppm + 10 μA from 10 to 100 mA
0.017% + 10 μA from 100 mA to 1 A
.04 A.C. ammeters
- At frequencies from 45 to 100 Hz
with least uncertainties of measurement of -
0.09% + 10 nA up to 100 μA
0.08% + 100 nA from 100 μA to 1 mA
0.08% + 1 μA from 1 to 10 mA
0.08% + 10 μA from 10 to 100 mA
0.1% + 100 μA from 100 mA to 1 A
0/07% + 1 mA from 1 to 11 A
.09 Ohmmeters
- 4 wire measurement
with least uncertainties of measurement of -
60 ppm + 2 mΩ up to 10 Ω
60 ppm + 10 mΩ from 10 to 100 Ω
60 ppm + 100 mΩ from 100 Ω to 1 kΩ
60 ppm + 1 Ω from 1 to 100 kΩ
60 ppm + 10 Ω from 100 kΩ to 1 MΩ
0.01% + 1 kΩ from 1 to 10 MΩ
0.05% + 100 kΩ from 10 to 100 MΩ
0.05% + 1 MΩ from 100 MΩ to 1 TΩ
3.12 Frequency and time measuring instruments and standards
- .11 Counters
- with least uncertainties of measurement of -
1 in 10⁵ + 1 count from 1 mHz to 50 MHz
.12 Time interval meters
- with least uncertainties of measurement of -
1 in 10⁵ up to 999 999 s
Accreditation No: 14283
Corporate Site No: 14276
(Scope Last Changed 29/07/03) |
|